The DYNEX product line, including the DSX® and DS2® instrument systems, has a proven track record of quality, innovation and reliability as a worldwide leader in fully automated systems and consumables.

DYNEX Agility® Specifications

Dimensions Metric Non Metric
Width: <1250 mm 50"
Depth: <900 mm 36"
Height: <1230 mm 49"
Footprint: <1200 mm x 650 mm 48" x 26"
Bench Weight: 213 kg (max.) 469 lbs (max.)
Ship Weight: 296 kg (max.) 650 lbs (max.)
Noise: Noise output <80 dB  
     
Power Requirements    
Voltage: 100 – 240V automatic switching  
Frequency: 50/60 Hz  
Power Consumption: Typically 400 VA (online UPS required)  

Reader Specifications

Photometric Range: 0.000 to 3.500 OD
Spectral Range: 405 nm to 690 nm
Precision: ±0.010 OD at 0.000 to 0.500, OD <1% CV at 0.501 to 2.000 OD, <1.5% CV at 2.001 to 2.500 OD
Accuracy: ±1% OD or 2.5% (0.000 to 3.000 OD) whichever is greater
Read Time: <30 seconds, single wavelength
<50 seconds, dual wavelength

Washer Specifications

Manifold Configuration: 8-way wash head
Programmable Volumes: 50-999 µl
Wash Containers: 4 wash bottles at 3.0 L, with quantitative level sensing
Clean Container: 1 wash bottle at 3.0 L, with quantitative level sensing
Waste Container: 10 L with quantitative level sensing
Residual Wash Volume: <3 µl per well with dual-axis sweep in a flat bottom well
Dispense Precision: ≤5% CV (with 300 µl in a 96 well plate)

Incubator Specifications

Number of Incubators: 12 (6 Ambient / 6 Elevated)
Temperature Range: RT + 4° C to 45° C (Elevated Incubator)
RT + 4° C (Ambient Incubator)
Temperature Accuracy: ± 1° C
Shaking: 14 Hz periodic or continuous

System Specifications

Number of Plates: 12 (additional continuous load capability)
Number of Sample Tubes: 200 (additional continuous load capability)
Number of Reagents: Up to 16 different SmartKit® Reagent Packs simultaneously (additional packs can be added using continuous load capability)
Number of Pipettes: 2 (sample + reagent)

Sample Pipetting Specifications

Sample Tip Size: 300 µl
Sample Pipetting Volume: 10-300 µl
Time to Dispense: 11 minutes (typical) 50 µl of 96 samples to plate from sample tubes or deep well plates
Sample Pipetting Precision: ≤3% CV (at any operating volume above 10 µl)
Sample Pipetting Accuracy: ± 2% of target volume
Dilution Range: 1 part in 199 one-stage dilution
1 part in 39,601 two-stage dilution
Number of Sample Tips: 20 racks of 112 tips (additional racks can be added using continuous load capability)
Sample Tube Dimensions: 10-17 mm diameter external dimension, 45-100 mm depth

Reagent Pipetting Specifications

Reagent Tip Size: 1200 µl
Reagent Pipetting Volume: 20-1200 µl
Number of Reagent Tips: 1 rack of 98 tips (additional racks can be added using continuous load capability)
Reagent Pipetting Precision: ≤3% CV at 10 shots at any volume in operating range
Reagent Pipetting Accuracy: ±2% of target volume (single shot mode)

Regulatory Compliance

Electromagnetic Perturbation: EMC Directive 2004/108/EC:
EN 61326-2-6:2006, IEC 61326-1:2006
Electrical Safety: BS EN 61010-1:2001, IEC 61010-1:2001, IEC 61010-2-101
Lead-free: All components are RoHS compliant
IVDD: CE marked per IVD 98/79/EC Directive


 